Abstract:After analyzing the suturing and knot tying operation of surgeons and the assignment of operation space, the motions of an expert surgeon are recorded in two projective views and analyzed quantitatively. The relationship between the measured data and robot dexterity is analyzed, which is taken as the basis for designing the manipulators. The accurate 3D dexterous operation space of the slave end-effector is obtained based on the projective geometry and micro-surgical operation data. According to the collected data of dexterous workspace, the manipulator parameters and movement ranges of the joints are computed using inverse kinematics principle. A new prototype robotic manipulator (MicroHand) for microsurgery has been produced based on theoretical analysis and design.
[1] Kwon D S,Woo K Y,Song S K, et al.Microsurgical telerobot system[A].Proceedings of the 1998 IEEE/RSJ International Conference on Intelligent Robots and Systems[C].Rictoria,Can:IEEE,1998,vol.2.945-950. [2] Ng W S,Davies B L.Robotic surgery [J].IEEE Engineering in Medicine and Biology,1993,12(1):120-125. [3] Kosuge K,Itoh T,Fukuda T.Scaled telemanipulation with communication time delay[A].Proceedings of the 1996 IEEE International Conference on Robotics and Automation [C]. Minneapolis, MN,USA: IEEE,1996,vol.3.2019-2024. [4] 陈梦东,王田苗,刘达,等.机器人辅助微损伤神经外科手术系统的研究及其临床应用[J].中国生物医学工程学报,2000,19(2):145-151. [5] Cavusoglu M C, Villanueva I, Tendick F.Workspace analysis of robotic manipulators for a teleoperated suturing task [A]. Proceedings of the 2001 IEEE/RSJ International Conference on Intelligent Robots and Systems[C].Maui,HI:IEEE,2001 vol.4.2234-2239. [6] 马香峰.工业机器人的操作机设计[M].北京:冶金工业出版社,1996. [7] 休斯敦,刘又午.多体系统动力学(上册)[M].天津:天津大学出版社,1987. [8] 冯友贤.血管外科学(第二版)[M].上海:上海科学技术出版社,1992.